Designing prototype compliant with European Regulation and FDA

A project log for Open assistive robotic arm for meal

An accessible robotic arm designed to help people with special feeding needs.

julien-oudinJulien OUDIN 08/31/2020 at 09:580 Comments

After the success of the first prototype which is a Proof of Concept, we decided working on a new prototype which meet Medical Device Regulation requirements, include security features to detect collision with the environment, review mechanical part which should be hermetic. Then we plan to do clinical trial with Christelle and other users.

Team

To reach the goal, we have been joined by 3 more students from different background.  

For the technical part we had hired two more student as intern from mechatronic background. Daria Kuznetsova, student in mechatronics engineering, who is working on actuator control and electronic design. Glenn Van Hemelryck, student in mechatronics engineering, who is working on software part, including user interface and features implementation. Georges Tawil, biomedical engineer, who is designing mechanical parts of the device. 

For regulation and quality part, we had hired, Ludovic Louisonne, student in biomedical engineering, who is working regulatory and quality strategy to provide a device compliant with European Regulation (Medical Device Regulation 745:2017, Quality Management System ISO 13 485:2016). He started a first work redefine device specifications according to the European medical device regulation. Second part, he was working on CE marking strategy via Biotika university company supervised by a regulatory expert.

Prototyping

After settling the work team, we define our roadmap. To develop quickly our next prototype, we make the choice to use actuator (motor, reducer, driver, encoder, torque sensor) on the market that it helps us to reduce the development time and skip actuator development. Later, we will thing to make our own actuator if we judge that can help us to reduce the cost of device.

Mechanical Design

First, we are working on the mechanical design. We reviewed our first prototype P0 designing then decide to add a 6th axis. We add an axis to adjust a lack of orientation which can be an issue in some case. We changed mechanical design of the arm to be more rigid, we use aluminium profile. We still use reducer for the firsts 3 axis for that we use timing belt. This kind of transmission is cheap, quiet and allowed to place actuators close to base of arm. Using timing belt allows to use actuator with smaller torque. We always think how to reduce the cost even if it’s can complicate prototype design. We also work on a user’s need, Christelle, our user ambassador, wants bring the device int the restaurant for that we thought on a design to be foldable.


Actuator selection

For selecting actuators, we defined requirements for each joint, angle range, maximal speed and the most important torque. The range was defined by planned trajectories to be functional. The maximal speed is defined very slow such as human motion. Finally, to determinate torque needed for each actuator we made static study for each joint considering all worst cases, particularly for the last joint we made test with food.

Software and Actuator controlled

First, we are working functional features:  

Then we will work on implementing security features to detect unusual collision with the environment to prevent from injury risks. that will allow us to reach safety requirement. 

Finally, we will make a user interface to facilitate utilisation, we thought to use a touchscreen on the device that increase the price of the device. Or we also thought to use Bluetooth connexion to connect with a smartphone application, it’s required for users to have a smartphone.


Medical Device Regulation

At the same time, we are editing European technical documentations, with help from a company and a test laboratory, these documentations consist of: 

We are following, the same process to get CE Marking which can guarantee design quality and security of our device. As medical device, it’s for us really important to provide and prove that our device is functional and can be used safely such as others devices on the market. 

Our goal it’s to have clear documentation on design part of technical documentations, there is another part which concern manufacturing we don’t approach for the moment.